Voyager DP-263-S Betriebsanweisung Seite 550

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 612
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 549
Appendix H Continuous Extraction Mode
H-22 PerSeptive Biosystems
H
Poor resolution in
Continuous Extraction
mode
(continued)
Beam guide wire
malfunction
Call PerSeptive
Biosystems Technical
Support.
Accelerating Voltage
malfunction
Call PerSeptive
Biosystems Technical
Support.
Microchannel plate
detector voltage set
incorrectly
Call PerSeptive
Biosystems Technical
Support.
Variable-voltage grid
contaminated with matrix
Call PerSeptive
Biosystems Technical
Support.
Poor mass accuracy in
Continuous Extraction
mode
Mass scale not accurately
calibrated
Recalibrate. See the
Data
Explorer Software User’s
Guide
, Section 5.3,
Manual Calibration.
Used different voltages
when acquiring sample
and calibration standard
Use same voltages when
acquiring sample and
calibration standards.
In Linear mode, used
different laser intensity
when acquiring sample
and calibration standard
Use same laser intensity
when acquiring sample
and calibration standards.
(continued on next page)
Incorrect masses entered
in calibration
Recalibrate. See the
Data
Explorer Software User’s
Guide
, Section 5.3,
Manual Calibration.
Table H-8 Spectrum Troubleshooting (Continued)
Symptom Possible Cause Action
Seitenansicht 549
1 2 ... 545 546 547 548 549 550 551 552 553 554 555 ... 611 612

Kommentare zu diesen Handbüchern

Keine Kommentare