Optimizing a Continuous Extraction Standard Instrument (.BIC) Setting
Voyager
™
Biospectrometry
™
Workstation User’s Guide H-5
H
Optimization
strategy
When optimizing a Continuous Extraction instrument settings
file, you:
• Start with a standard instrument settings file
• Fine-tune laser setting for major improvement in
performance
• Optionally adjust Grid Voltage% and Guide Wire
Voltage% for slight improvement in performance
Optimizing
1. Open a standard instrument settings file for the mass
range you are analyzing.
If an instrument settings file for the mass range you are
analyzing does not exist, open a standard instrument
settings file with the closest higher mass.
2. Save the instrument settings file under a new name.
Standard instrument settings files are read-only and
cannot be saved.
3. Select a calibration (.CAL) file. If you are screening
samples, use the default calibration.
4. If the mass range needs adjusting, change the digitizer
setting. See Section 5.3.5, Understanding Digitizer
Settings.
5. To include matrix peaks in the spectrum for calibration,
deselect the Low Mass Gate field, or set the Starting Mass
to a mass below the matrix peak mass.