Voyager DP-263-S Betriebsanweisung Seite 397

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 612
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 396
Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-85
6
The number and placement of standards needed depend on
your application. See “Mass calibration standards” on
page 6-83.
Hint: You can use the same calibration file more than once
in a sequence run.
Sample and
standard in the
same sample
position
You may see improved mass accuracy by spotting sample and
standard in as close together as possible within a sample
position (Figure 6-25). A standard spot within 1 mm of a
sample spot can yield mass accuracy close to that expected
for internal calibration.
NOTE: Use plates without laser-etched sample positions or
wells if you spot standard and sample as shown below.
Figure 6-25 Location of Sample and External Standard
for Maximum Mass Accuracy
Sample
Standard
Logical
coordinates (microns)
Seitenansicht 396
1 2 ... 392 393 394 395 396 397 398 399 400 401 402 ... 611 612

Kommentare zu diesen Handbüchern

Keine Kommentare