Startup and Shutdown
Voyager
Biospectrometry
Workstation User’s Guide 2-47
2
If your system includes a LeCroy digitizer, wait
approximately one minute until the digitizer completes
its internal calibration before starting the Instrument
Control Panel.
4. Press the EMIS button on the Vacuum Gauge Panel to
turn on the gauges. Observe the vacuum gauge, and
allow the pressure on BA1 and BA2 to reach:
Press Chan to cycle between BA1 and BA2.
NOTE: After shipment or venting the system, it will
require one to two days after startup to reach this
vacuum level.
5. Log on to the Voyager Workstation using your User Name
and Password. See your system administrator for your
User name and Password.
6. Start the Voyager Instrument Control Panel software by
double-clicking the Voyager Instrument Control icon on
the Windows NT desktop.
Model
BA1 Pressure
(Torr)
BA2 Pressure
(Torr)
Voyager-DE and
Voyager-DE PRO
Less than 10
-6
Less than 10
-6
Voyager-DE STR Less than 5x10
-7
Less than 5 x10