Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-73
6
Calibration File
(required entry
for External,
Internal-Update
calibration,
optional entry
for Internal
calibration)
.CAL file to use to calibrate the data file acquired in this row. If
you specify External or Internal-Update, a .CAL file name is
required, even if a corresponding .CAL file with constants does
not exist (see below for further explanation).
You can specify an existing .CAL file or the name of a .CAL file
that will be created by a previous line in the Sequence.
The way the .CAL file is used depends on the calibration type:
External—Constants applied.
Internal—If .CAL file specified, constants applied.
Internal-update—If the .CAL file specified exists, constants
are applied before calibration, then the .CAL file is
updated
with new constants after calibration.
If the .CAL file specified does not exist, no constants are
applied before calibration, the .CAL file is
created
containing
new constants after calibration.
NOTE: If the
.
CAL file specified for Internal-Update does not
exist, the software creates the .CAL file during processing in the
path you specify. If you do not specify a path, the software
creates the .CAL file in the C:\VOYAGER directory.
NOTE: If the .BIC specifies a .CAL file, the .CAL in the .BIC is not
updated, unless it is the same .CAL specified in the Sequence
Control Panel.
For information on how .CAL files are used in Sequence Control
Panel, see Section 6.7.1, Understanding Settings,
Macros, and Calibration.
Table 6-4 Run List Parameters (Continued)
Field Description