Chapter 6 Acquiring Mass Spectra
6-86 PerSeptive Biosystems
6
To acquire sample and standard, create two search pattern
files to analyze the sample and standard spots. See “Creating
search pattern files for close external calibration on separate
spots” on page 6-86, for more information. Create two
instruments settings (.BIC) files that contain the same settings,
but specify the sample search pattern file and the standard
search pattern file.
To run samples spotted as illustrated in Figure 6-25 (assuming
that subsequent positions are spotted in the same way), set up
the run list as shown in the following table.
Creating search
pattern files for
close external
calibration on
separate spots
To create search pattern files for close external calibration on
separate spots:
1. Spot sample and standard as close to each other as
possible (within 1 mm) (see Figure 6-25 on page 6-85).
NOTE: For optimum mass accuracy, calibration
standards must be as close to unknowns as possible.
An internal standard spot within 1 mm of a sample spot
can yield mass accuracy close to the expected mass
accuracy for internal standard calibration on your
system.
Position Base File
Name
Instrument Settings File Calibration Type Calibration
File
91 STD1 STD.BIC (specifies STD.SP) INTERNAL-UPDATE STD.CAL
91 SAMP1 SAMP.BIC (specifies SAMP.SP) EXTERNAL STD.CAL
92 STD2 STD.BIC (specifies STD.SP) INTERNAL-UPDATE STD.CAL
92 SAMP2 SAMP.BIC (specifies SAMP.SP) EXTERNAL STD.CAL
93 STD3 STD.BIC (specifies STD.SP) INTERNAL-UPDATE STD.CAL
93... SAMP3 SAMP.BIC (specifies SAMP.SP) EXTERNAL STD.CAL