Chapter 3 Preparing Samples
3-44 PerSeptive Biosystems
3
5. Select a Plate ID. The .PLT file and alignment information
associated with the Plate ID are automatically loaded.
Alternatively you can specify a new Plate ID and select
a .PLT file. For more information, see “Assigning Plate
IDs” on page 3-39.
6. Click OK.
The sample plate is loaded and aligned as needed. For
more information, see “How the system aligns a plate”
on page 2-40.
It takes a minute or two for the sample plate to reach
the correct position. While the sample plate is moving,
the Load/Eject Status dialog box displays messages
about the status of the hardware.
Voyager-DE STR
To load sample plates:
1. Eject the sample holder as described in Section 3.4.1,
Ejecting the Sample Holder.
2. Hold the sample plate vertically, with the sample surface
facing to the right, and with the slanted underside of the
plate facing toward the back of the instrument.
3. Slide the sample plate into the holder from the front until it
snaps into place (Figure 3-7). The ball bearings on the
holder snap into the plate.
CAUTION
If the sample plate does not snap into place, it may be
inserted into the holder the wrong way, and it may jam
inside the instrument. Remove the plate, slide it into the
holder with the slanted underside of the plate facing to
the left and toward the back of the instrument, and snap
it into place.