Chapter 3 Preparing Samples
3-20 PerSeptive Biosystems
3
3.1.5.1 Washing
When to use
Use this technique when you know the solubility of the
contaminant. You can wash a dried sample directly on the
sample plate.
What you need
Select a solvent in which the contaminant is more soluble than
the matrix and the sample. For example, use:
Cold water (to prevent sample dissolving) with 0.1% TFA
for a polar contaminant, such as a salt or buffer
5% isopropanol in water for a non-ionic detergent such as
octyl glucoside
Procedure
To wash:
1. Apply sample and matrix to the sample plate and allow
to dry.
2. Add 1 to 3 µl of solvent to the dry sample/matrix.
3. Wait 10 seconds, then remove the liquid.
NOTE: Try not to touch the sample plate with the pipet
tip. If you do touch the spot, do not analyze the area
you touched.
4. Repeat step 2 and step 3 one more time.
5. Allow the sample plate to dry before analyzing.
If one wash does not improve results, you can wash
again.