
Troubleshooting
Voyager
™
Biospectrometry
™
Workstation User’s Guide 8-25
8
E09 error message
displayed on gauge
controller for BA1 or BA2
BA1 or BA2 gauges shut
down due to high
pressure. May be caused
by:
• Sudden increase in
pressure when
sample plate inserted
• Vacuum leak
Press the EMIS button on
Pressure Gauge Control
Panel to turn off. Press
again to turn on.
Call PerSeptive
Biosystems Technical
Support if error occurs
again.
E08 error message
displayed on gauge
controller
Vacuum gauge board
failed
Power down the mass
spectrometer and then
power up.
Call PerSeptive
Biosystems Technical
Support if error occurs
again.
Table 8-5 Vacuum Gauge Panel Troubleshooting
Symptom Possible Cause Action
Kommentare zu diesen Handbüchern