Acquiring Spectra from the Sequence Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-89
6
2. Load the sample plate, select the position containing
sample and standard, and display the Sample View.
3. Note the relative x and y coordinates for a minimum of
three positions on the standard spot and a minimum of six
positions on the sample spot.
4. Create the search pattern file using Windows Notepad as
described “Creating a search pattern file” on page 6-47.
Figure 6-26 shows an example search pattern (.SP) file
for internal calibration using separate spots.
Figure 6-26 Example .SP File for Internal Calibration
Using Separate Spots
;internal calibration using two spots, 5/25/99
-1048,635
-857,635
-762,635
;sample spots
;std spots
635,-635
730,-635
825,-635
635,-730
730,-730
825,-730