Chapter 5 Optimizing Instrument Settings
5-52 PerSeptive Biosystems
5
When to increase
Increase the Vertical Scale if signal goes offscale
(Figure 5-12).The Vertical Scale setting is typically between
50 mV and 1,000 mV. If you set it at 1,000 mV (the maximum
Vertical Scale setting) and signal is offscale, decrease the
laser intensity to decrease the signal intensity. The offscale
signal exceeds 64,000 counts (see the right hand axis).
When to decrease
Decrease the Vertical Scale if signal intensity of tallest peak is
not greater than approximately 20 percent of the vertical
maximum (Figure 5-12). If you are at the minimum Vertical
Scale setting and signal is not greater than approximately
20 percent of the vertical maximum, follow the guidelines in
Section 5.4.4, Optimizing Signal-to-Noise Ratio, to increase
signal intensity.
Vertical Offset
The Vertical Offset allows fine adjustment of the spectrum
baseline position. It is specified as a percentage of the Vertical
Scale.
For most applications, leave the Vertical Offset at 0%.
Figure 5-13 illustrates the effect of setting a positive or
negative offset. A positive offset shifts the baseline up. A
negative offset shifts the baseline down.
NOTE: If you set a Negative Vertical Offset, some
low-intensity noise signal may not be collected in the data
file. Some processing functions, for example Baseline
Correction or Noise Filter, require complete collection of
noise signal to generate valid results.
Figure 5-13 Effect of Vertical Offset in Digitizer Settings
Negative
Offset
Vertical
Positive
Offset
Vertical
Original data
scale (0 Vertical
noise signal below
Offset)