Chapter 5 Optimizing Instrument Settings
5-38 PerSeptive Biosystems
5
Table 5-11 Automatic Control Parameters—Sample Positioning
Parameter Description
Automated Sample Positioning
Use Automated
Sample Positioning
Enables or disables automated sample positioning.
For more information, see “Sample positioning” on
page 6-39.
Search Pattern File Determines the search pattern used when Use Automated
Sample Positioning is enabled. See Section 6.6.4, Search
Pattern Files, for more information.
Number of Positions Displays the number of positions in the currently selected
search pattern file.
Stop Conditions Enable Stop Conditions if you want acquisition to stop
after:
• Specified consecutive number of acquisitions fail
acceptance criteria
• Specified consecutive number of laser adjustments
do not yield acceptable data