Voyager DP-263-S Betriebsanweisung Seite 286

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 612
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 285
Chapter 5 Optimizing Instrument Settings
5-56 PerSeptive Biosystems
5
Strategy
To optimize instrument settings, do the following:
1. Open a .BIC file for the mass range you are analyzing.
For information on mass ranges in .BIC files, see
Section 5.1.2, Standard Instrument
Settings (.BIC) Files Provided.
If a .BIC file for the mass range you are analyzing does
not exist, open a standard .BIC file with the closest
higher mass.
Hint: To optimize a wide mass range, select a .BIC file
with the mass of the highest component in the range.
2. Find an adequate laser setting. See Section 5.4.2,
Determining the Laser Setting.
3. Optimize resolution. See Section 5.4.3, Optimizing
Resolution.
4. Optimize signal-to-noise ratio. See Section 5.4.4,
Optimizing Signal-to-Noise Ratio.
Remaining parameters in the standard instrument settings
provided are optimized for your workstation and should not
require adjustment.
For more information on the remaining instrument settings
parameters, see Section 5.2, Instrument Settings
Parameter Descriptions.
Seitenansicht 285
1 2 ... 281 282 283 284 285 286 287 288 289 290 291 ... 611 612

Kommentare zu diesen Handbüchern

Keine Kommentare