Chapter 7 PSD Analysis
7-34 PerSeptive Biosystems
7
During
acquisition
The following occurs:
Mass range for the segment is set to:
Acquisition starts.
All instrument settings except Shots/Spectrum are
disabled. Settings remain inactive until you stop
the PSD experiment after you acquire all
segments of interest.
Acquisition of the segment continues until the
number of Shots/Spectrum specified in Spectrum
Acquisition on the Instrument Settings control
page is collected, or until you select Stop
Acquisition from the Acquisition menu.
The spectrum is displayed in the Current Trace in
the Spectrum window.
Changing
settings
During acquisition, you can vary laser intensity,
parameters in the segment list, and Shots/Spectrum.
Mass Equivalent to
Start (Precursor mass/4)
which is equal to
(Precursor flight time/2)
End (Mirror to Accelerating Voltage Ratio
2
x
Precursor mass)
which is equal to
(Mirror to Accelerating Voltage Ratio x
Precursor flight time)