Voyager DP-263-S Betriebsanweisung Seite 279

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 612
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 278
Impact of Changing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-49
5
Effects of
adjusting Bin size
A smaller Bin size (lower number of nanoseconds):
Increases resolution between peaks when peaks are
narrow in time width (see Figure 5-11).
Increases the size of the data file because the mass
range is recorded with finer sampling and therefore
increases the number of data points collected.
Figure 5-11 Effect of Bin Size (nsec) on Peak
Resolution and Data Quality
A larger Bin size (higher number of nanoseconds):
Decreases resolution between peaks when peaks are
narrow in time width (see Figure 5-11).
Decreases the size of the data file because the mass
range is recorded with coarser sampling (decreases the
number of data points collected).
Data collected at
2 nanoseconds
Data collected at
8 nanoseconds
Baseline resolution
between 2 adjacent
peaks
Partial resolution
between same
peaks
Seitenansicht 278
1 2 ... 274 275 276 277 278 279 280 281 282 283 284 ... 611 612

Kommentare zu diesen Handbüchern

Keine Kommentare