Acquiring in Manual Mode from the Instrument Control Panel
Voyager
Biospectrometry
Workstation User’s Guide 6-19
6
6.2.2 Manually Accumulating Spectra
from Multiple Acquisitions
You can manually accumulate spectra from different
acquisitions to improve resolution and signal-to-noise ratio of
your data.
Accumulating
To manually accumulate spectra from multiple acquisitions:
1. Acquire a spectrum and evaluate the data as described
in Section 6.2.1, Manually Acquiring,
Evaluating, and Saving Spectra.
2. When acquisition is complete, do not save the data.
3. If the spectrum is acceptable, select Accumulate
Spectrum from the Acquisition menu or click to
accumulate the spectrum.
The software adds an Accumulated trace to the
Spectrum window and copies the Current trace in the
Accumulated trace.
If the spectrum is not acceptable, start a new
acquisition. The Current trace will be overwritten by the
new acquisition.
CAUTION
If you change any instrument settings, with the
exception of Shots/Spectrum and laser intensity, the
Accumulated trace is automatically cleared before the
next acquisition.
4. Repeat step 1 though step 3 to acquire and accumulate
additional spectra. Each spectrum you accumulate is
averaged with the existing Accumulated trace.
NOTE: If needed, select Clear Accumulated trace
from the Acquisition menu or click .
5. When the data is acceptable, select (click on) the
Accumulated trace.
6. Select Save Spectrum from the Acquisition menu.