Appendix H Continuous Extraction Mode
H-16 PerSeptive Biosystems
H
H.2.2.3 Verifying Threshold Setting
Move to a new area of the sample well that contains the same
sample. Acquire a spectrum to verify that the setting is valid
for the laser power selected, and not caused by:
Sample surface excitement caused by the previous
higher laser power setting
Sample consumption
If the setting is valid, you see a spectrum similar to the one just
acquired. You may need to adjust the laser slightly (5 to
10 counts). Examine several spots in the sample well to make
sure you have not tuned the laser intensity on a “cold spot”.
You can further fine-tune the laser setting by setting a smaller
laser step size in Hardware Configuration. For example,
instead of a step size of 10, change the step size to 2.
Make sure peak shape and resolution are acceptable. See
Section H.2.3, Checking Resolution. If you do not see a signal
at the current setting, increase the laser power.
NOTE: If peak intensity of the final scan is near maximum
intensity, it is very possible that some scans averaged into
the final scan are saturated. Reacquire using a lower laser
power.