Seitenansicht 254
Instrument Settings Parameter Descriptions
Voyager
Biospectrometry
Workstation User’s Guide 5-25
5
Instrument Mode
parameters
Click the Instrument Mode tab to display the Instrument
Mode page (Figure 5-4).
Instrument Mode parameters are described in Table 5-6.
Table 5-6 Instrument Mode Parameters
Parameter Description
Operation Mode Specifies Operation Mode:
Linear
Reflector
•PSD
NOTE: Reflector and PSD modes are not available on the
Voyager-DE system. If you have a Voyager-DE system,
buttons for these two modes are not displayed. For
information on PSD parameters see, Chapter 7, PSD
Analysis.
Extraction Type Specifies type of extraction to use:
Delayed—Described in Section 1.4, Voyager-DE™
(Delayed Extraction™) Technology.
Continuous—For more information, see Appendix H,
Continuous Extraction Mode.
Polarity Type Specifies polarity of spectra to acquire:
Positive
Negative
Laser Type Specifies type of laser installed:
Internal
External (optional)
Seitenansicht 254